Analyzer Messages

"No Configuration File Loaded"

"No Configuration File Loaded"
This is usually caused by trying to load a configuration file for one type of
module/system into a different type of module/system.
Verify that the appropriate module has been selected from the Load
{module} from File {filename} in the HP 16500A/B/C disk operation
menu. Selecting Load {All} will cause incorrect operation when loading
most analysis probe configuration files.
See Also
Chapter 2 describes how to load configuration files.

"Selected File is Incompatible"

This occurs when you try to load a configuration file for the wrong module.
Ensure that you are loading the appropriate configuration file for your logic
analyzer.

"Slow or Missing Clock"

This error message might occur if the logic analyzer cards are not firmly
seated in the logic analysis system mainframe. Ensure that the cards are
firmly seated.
This error might occur if the target system is not running properly.
Ensure that the target system is on and operating properly.
If the error message persists, check that the logic analyzer pods are
connected to the proper connectors on the analysis probe. See Chapter 2
to determine the proper connections.
5–14
HP E2416B 80196 Analysis Probe