No Activity On Activity Indicators; No Trace List Display; Analyzer Won't Power Up - HP E2416B User Manual

Analysis probe for intel 80196
Table of Contents

Advertisement

Analyzer Problems

No activity on activity indicators

No activity on activity indicators
Check for loose cables, board connections, and analysis probe
connections.
Check for bent or damaged pins on the analysis probe.

No trace list display

If there is no trace list display, it may be that your trigger specification is not
correct for the data you want to capture, or that the trace memory is only
partially filled.
Check your trigger sequencer specification to ensure that it will capture
the events of interest.
Try stopping the analyzer; if the trace list is partially filled, this should
display the contents of trace memory.

Analyzer won't power up

If the logic analyzer power is powered down when it is connected to a
powered-up target system, the logic analyzer may not be able to power up.
Some logic analyzers are inhibited from powering up when they are
connected to a target system that is already powered up.
Disconnect all logic analyzer cabling from the analysis probe. This will
allow the logic analyzer to power up. Reconnect logic analyzer cabling
after power up.
5–4
HP E2416B 80196 Analysis Probe

Advertisement

Table of Contents
loading

Table of Contents