Quality measurement

4.8.3
The Spine 18 is positioned on the patient table.
(1) Position of the phantom holder: Shoulder Small 6
(2) Position of the phantom holder: Shoulder Large 6
1
Position the phantom holder 25 on the patient table.
2
Position the 5300 ml plastic bottle on the phantom holder so that
the bulge of the bottle is facing the magnet tunnel.
3
Position the coil on the plastic bottle, so that the bulge of the
plastic bottle fits into the coil cup.
4
Connect the coil plug to sockets 1 or 2.
5
Align the laser light localizer with the center marking of the coil.
6
Start the quality measurement.
assurance (software))
4.9

Wrist 8

4.9.1

Description

The Wrist 8 is used to examine wrists, hands, and fingers. The coil is
used with the patient in the Head First (prone position).
Accessories
(
Page 77 Coil accessories)
The coil consists of the following components: