Coils
Table of Contents

Advertisement

Configuration
MAGNETOM Sempra
Print No. M4-08006G.623.01.02.02
Quality measurement
4.10.3
The Spine 18 or the table pads are positioned on the patient table.
1
Position the lower part of the Extremity 12 on the patient table.
2
Position the phantom holder 28 in the lower part of the coil and
position the plastic bottle 1900 ml on the phantom holder so that
the bottom of the bottle is facing the magnet bore.
3
Position the upper part of the coil on the lower part.
4
Connect the coil plug to coil socket 1 or 2.
5
Align the laser light localizer with the center marking of the coil.
6
Start the quality measurement
assurance (software)).
4.11

Foot/Ankle 10

Description

4.11.1
The Foot/Ankle 10 is used to examine feet and ankles.
Accessories
(
The coil consists of the following components.
(1) Handgrip
(2) Upper part
(3) Center marking
(4) Tilt support mechanism
(5) Base plate

Use

4.11.2
A sample examination of the right foot/ankle is described.
(