Use - Siemens Healthineers MAGNETOM Sempra Operator's Manual

Coils
Table of Contents

Advertisement

4
Coils
General information
Examining the shoulder and
68

Use

4.13.2
In what follows, two examination examples are described:
Shoulder and hip in supine position (Flex Large 4)
Hand and elbow in prone and supine positions (Flex Small 4)
1
Use the Flex Coil Interface as described.
2
Position the coil with the flat side against the region of interest.
Examinations of the shoulder and hip may show strong artifacts
caused by layers of fat.
3
To avoid artifacts, increase the distance between the coil and the
area to be examined by using the iPAT cushion.
Ensure that the coil cable does not hang over the sides of the
patient table.
hip
The Spine 18 is positioned on the patient table.
The Flex Coil Interface is connected.
The coil is connected at the Flex Coil Interface.
1
Attach the strap to the side of the patient table opposite the
shoulder to be examined.
2
Position the coil.
3
Ensure that the region to be examined is contained within the side
flaps of the coil.
4
Secure the coil with the strap.
Coils | Operator Manual
Print No. M4-08006G.623.01.02.02