Spine; Description - Siemens Healthineers MAGNETOM Sempra Operator's Manual

Coils
Table of Contents

Advertisement

4
Coils
Configuration
44
2
Position the 5300 ml plastic bottle on phantom holder 29 so that
the bulge of the bottle is positioned as far as possible in the lower
part of the coil.
3
Place the upper part on the lower part of the coil and slide it
forward until it locks into place.
4
Align the laser light localizer with the corresponding center
marking of the coil.
5
Start the quality measurement.
assurance (software))
4.4

Spine 18

4.4.1

Description

The Spine 18 is used for examining the spine and for combined
measurements with other coils.
Up to four clusters (groups of coil elements) of the Spine 18 can be
used at a time.
Accessories
(
Page 77 Coil accessories)
The coil consists of the following components:
(
Page 33 Performing quality
Coils | Operator Manual
Print No. M4-08006G.623.01.02.02