Coils
Performing the measurement
54
5
When positioning the patient with the arms above the head: Use
the insulation pad to avoid current loops.
6
Use suitable positioning aids.
1
Align the laser light localizer with the center marking of the coil.
2
Start the measurement.

Quality measurement

4.6.3
All RF coils and cushions are removed from the patient table.
The Breast 4 is positioned at the head end of the patient table.
The coil plug is connected to coil socket 1.
1
Position plastic bottle 1900 ml into the left coil element flush with
the inside curved edge.
2
Align the laser light localizer with the center marking of the coil.
3
Start the quality measurement
assurance (software)).
4
Position plastic bottle 1900 ml into the right coil element flush
with the inside curved edge.
5
Align the laser light localizer with the center marking of the coil.
6
Start the quality measurement
assurance (software)).
(
Page 33 Performing quality
(
Page 33 Performing quality
Coils | Operator Manual
Print No. M4-08006G.623.01.02.02