Use - Siemens Healthineers MAGNETOM Sempra Operator's Manual

Coils
Table of Contents

Advertisement

Positioning the coil
Positioning the patient
MAGNETOM Sempra
Print No. M4-08006G.623.01.02.02

Use

4.9.2
In the following section, the examination of the right hand is
described. The examination procedure of the left hand is analog to
the following description - unless the coil is positioned on the
opposite side of the patient table.
The Spine 18 is positioned on the patient table.
1
Place the Wrist 8 on the patient table.
2
Center the coil as much as possible for optimal image.
3
Connect the coil plug to coil socket 1 or 2.
4
Position the wrist pad, palm pad and upper pad as required in the
coil.
(
Page 57 Description)
1
Use suitable positioning aids.
2
Position the patient in the prone position with the head toward the
magnet.
3
Open the upper part of the coil with the unlocking mechanism.
4
Position the patient's hand in the coil and ensure that the pads are
positioned correctly.
5
If necessary, use the sandbags to stabilize the coil.
In the horizontal coil position, you can also turn the coil at an
angle to adjust it to the patient's position.
6
Close the upper part of the coil. It has to audibly lock into place.
7
Stabilize the fingers with the suitable wedge cushions.
The coil will function correctly only if the upper part is fully
closed.