To adjust the position of the straps along the groove, grasp the
mounting device of the strap and slide it to the required position.
Position the mounting device so that the strap is vertical. If the
strap is positioned at an angle, this may cause the mounting device
to shift within the groove.
3
To remove the strap: Press against the middle part of the mounting
device.
3.3

Positioning without the Head/Neck 16

If you do not use the Head/Neck 16, the head of the patient is
positioned in the back-of-the-head support.
The back-of-the-head support must be positioned correctly so that the
system knows the exact patient position and can determine the
correct SAR values.
 CAUTION
Patient positioning in Head First orientation without a head or
head/neck coil!
Wrong SAR calculation resulting in local overheating
Always use the back-of-the-head support if no head or head/
neck coil is used.
Make sure that the back-of-the-head support is positioned
correctly on the head-heel insert so that it is flush with the
edge marking of the head-heel insert.