Frequently Asked Questions:
1. Rotate the Trim Knob to highlight the legend for MHR/P. (This legend is located
approximately in the center of the display, and may indicate MECG, Pulse or INOP,
depending on the settings that are currently enabled.
2. Once the MHR/P legend is highlighted, press the Trim Knob. The display changes to
show the MHR/P Setup screen.
3. Now rotate the Trim Knob to select (highlight) the Alarm Volume: setting located
immediately to the right. The setting is in the range of 1 to 9.
4. Once the Alarm Volume: setting is highlighted, press the Trim Knob again. The
current setting displays in blinking inverse video, and an audio tone, indicative of the
alarm volume, is emitted from the speaker.
5. Now rotate the Trim Knob to change the current setting as desired. With each setting
change, an audio tone, indicative of the alarm volume, will be emitted from the
speaker.
6. Once you set the desired alarm value, press the Trim Knob to confirm your selection.
The current value setting stops blinking. A final audio tone, indicative of the alarm
volume, is emitted from the speaker.
7. Now rotate the Trim Knob to select (highlight) the Exit item on the bottom menu.
8. Press the Trim Knob again to return to the main monitoring display screen.
1. Rotate the Trim Knob to highlight the legend for MSpO
slightly above center, on the right side of the display.)
2. Once the MSpO
legend is highlighted, press the Trim Knob. The display changes to
2
show the MSpO
Setup screen.
2
3. Now rotate the Trim Knob to select (highlight) the Alarm Volume: setting which is
located immediately to the right. The setting is in the range of 1 to 9.
4. Once the Alarm Volume: setting is highlighted, press the Trim Knob again. The
current setting displays in blinking inverse video, and an audio tone, indicative of the
alarm volume, is emitted from the speaker.
5. Now rotate the Trim Knob to change the current setting as desired. With each setting
change, an audio tone, indicative of the alarm volume, will be emitted from the
speaker.
6. Once you set the desired alarm value, press the Trim Knob to confirm your selection.
The current value setting stops blinking. A final audio tone, indicative of the alarm
volume, is emitted from the speaker.
7. Now rotate the Trim Knob to select (highlight) the Exit item on the bottom menu.
8. Press the Trim Knob again to return to the main monitoring display screen.
250cx Series Maternal/Fetal Monitor
2036946-001