MHR/P Setup Screen

Auto
MECG
MSpO
2
On
Off
0
1
2
3
4
1
5
2
6
3
7
4
8
5
9
6
7
8
9

Source

7-4
Maternal Heart/Pulse Rate Monitoring: MHR/P Setup Screen
Select the mode title softkey—MECG or Pulse—to access the MHR/P Setup screen.
(Refer to the following figure.)
NOTES
The figure below provides an example of MECG selected as the MHR/P
source, as indicated by the MECG mode title. When either MSpO
are selected as the MHR/P source, the mode title changes to Pulse.
The Lead source and Pacer fields apply to MECG only.
FECG
US
165 172
MHR/P Setup
MECG
Source:
Auto
HR/PR Trace:
Off
Volume:
5
Alarms
High
MHR/P:
Alarm Volume
100-250
This field selects the MHR/P source. When Auto is selected, the monitor checks for
parameter availability and use in the following order: MECG, MSpO
a source is not available, the next available source is automatically selected.
IMPORTANT
WAVEFORM—The MHR/P Source field is independent of the
waveform selected on the normal operating screen. For example,
you can select MECG as the MHR/P source yet display the
MSpO
plethysmograph waveform. Or, you can select MSpO
2
the source and display MECG as the waveform.
250cx Series Maternal/Fetal Monitor
2036946-001
TOCO
30
MECG
89
Lead:
II
Pacer:
Off
Low
120
50
bpm
5
or NIBP
2
I, II, III
On
Off
35-120
Exit
MHR/P Setup Screen
, then NIBP. If
2
as
2
Revision C