ALERT PARAMETERS SUMMARY ON FHR SETUP SCREEN
Parameter
Column 1
Good
Moderate
Poor
Unknown
when UC is the only
active parameter)
Poor
Poor
Signal Quality
Poor
Poor
Poor
no data
Revision C
Spectra Alerts: Alert Parameters
Summary of Alert Parameters
Column 2
(appears
With UC
With UC
With UC
250cx Series Maternal/Fetal Monitor
2036946-001
TREND SCREEN
Message
Alert pending.
Alert has not been silenced.
SIGNAL QUALITY?
Three minutes of unsatisfactory data
(FECG).
Five minutes of unsatisfactory data
(ultrasound).
SIGNAL QUALITY?
Alert silenced.
The alert is deleted after 3 minutes of
satisfactory data.
Data unsatisfactory with uterine contractions
present.
SIGNAL QUALITY?
Alert has not been silenced.
Data between uterine contractions is
acceptable; data during contractions is
either poor quality or absent. Alert occurs
after 1, 2, or 3 uterine contractions
depending on what events preceded it.
SIGNAL QUALITY?
Alert silenced.
The alert is deleted following two uterine
contractions with satisfactory data or 10
minutes of "good" data.
Message appears if there is a problem with
the monitor or the Spectra Alerts option.
REPAIR
Contact Biomedical Department or Service
Representative.
Criteria