Name
E
Mark [Offset] Button
F
UA Reference Button
G
Paper Advance Button
H
Record Button
I
Power Indicator
J
Record Indicator
K
Light Button
L
Recorder Door Latch
M
Power Switch
N
Strip Chart Recorder
O
MATERNAL NIBP Connector
P
MATERNAL SpO
2
Q
FECG/MECG Connector
3-4
Controls, Indicators, and Connectors: Front Panel Description
The Mark [Offset] button is a multi-function button.
Mark: Pressing this button prints an event mark
bottom two lines of the top grid).
Offset: When the Heart Rate Offset mode is enabled, pressing and holding this
button shifts the secondary FHR trend +20 bpm for visibility purposes. Refer to
"Fetal Heart Rate Offset" on page 5-11.
The UA Reference button sets a baseline for uterine activity pressure monitoring.
Refer to Chapter 6, "Uterine Activity Monitoring".
Pressing this button advances chart paper at a rate of 40 cm/min for as long as the
button is held down.
The Record button selects one of three recorder states: on, maternal-only mode, or
off. Refer to Chapter 11, "Recorder Modes". Factory default is OFF.
The indicator lights green when the monitor is turned on.
Indicator Status
on
off
three short flashes
every 5 sec
flashes on and off
Illuminates the strip chart paper for night time visibility. Factory default is ON.
Opens the strip chart recorder door to add, remove, or adjust the paper.
Moving the switch to the on position (I) turns the monitor on; moving the switch to the
off position (O) turns the monitor off.
Annotations and trends are printed on the strip chart paper. Two paper styles are
available. Refer to Chapter 4, "Setup Procedures", for instructions on loading strip
chart paper into the recorder. Refer to Chapter 11, "Recorder Modes" for additional
information about trends and annotations.
Connect a pneumatic hose and blood pressure cuff assembly to this black twin lumen
receptacle.
Connector
Connect a 250cx Series MSpO
only Nellcor Maternal Oxygen Saturation Sensors if Nellcor technology is installed in
your monitor, Masimo Sensors if Masimo technology is installed in your monitor, or
Ohmeda Sensors if Ohmeda technology is installed in the monitor.
Connect an FECG cable/legplate or MECG cable plug to the FECG/MECG receptacle.
Cables with rectangular plugs connect directly to the FECG/MECG receptacle. Cables
with round plugs require an FECG/MECG adapter. Refer to "MECG Ordering
Information" on page 18-5 for the adapter part number. This adapter is used for dual
ECG monitoring as well. The adapter branches into two cables, each with a round
receptacle at the end: one branch is labeled MECG; the other branch is labeled FECG.
250cx Series Maternal/Fetal Monitor
Front Panel
Description
Recorder Status
on
off
maternal-only mode
error condition
intermediate cable to this royal blue receptacle. Use
2
2036946-001
on strip chart paper (on the
Revision C

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Corometrics 256cxCorometrics 259cx