Operator's Manual Simple AFG Application

Trigger Settings tab

The Trigger In settings parameters are located in the Trigger Settings Tab. These parameters are shared
by all channels of the instrument.
Trigger In Setting
Source
Timer Interval [s]
Threshold [V]
Edge
Description
Trigger Button: the Trigger event is provided to the instrument by pressing
the Trigger button on the keyboard or the Trigger button on the menu toolbar
or by issuing a trigger by Remote Command.
Timer: the Trigger is internally generated by a Timer. The Timer count interval
is set by Interval [s] textbox.
Trigger In Connector: a Trigger event is generated by the signal applied
externally to the "Trigger In" BNC connector when it crosses the selected
Threshold.
Note: the Trigger buttons and the Trigger from remote command are always
active, independently from the selected Trigger Source.
Set the timer count interval. It has effect only when the selecting trigger
Source is Timer. The Interval range is from 13.4ns to 100s. The edited value is
automatically rounded to the closest value that the hardware can implement.
It is the threshold that the external signal applied to the "Trigger In" BNC
connector must cross to issue a Trigger event to the instrument. It has effect
only when the selected Source is Trigger In Connector.
When Rising Edge is selected the trigger is detected when the signal on the
"Trigger In" BNC connector crosses the threshold from low to high. The Falling
71