Operating the System
Wallstand Examination Procedures
4.9.1.3 Basic Flow
Fig. 4-31 Wall Flexible mode display
• The SID is displayed.
Activate the mode by pressing the servo button.
• All movements are available.
• The overhead tube crane moves automatically to its programmed position.
• The overhead tube crane either goes direct to an aligned position to the detector or waits
in the transportation height until detector is moved (wait or no wait selected).
• The mode stays activated. The overhead tube crane beeps once and the servo mode
indication light flashes.
Adjust the height of the wallstand.
Move the detector holder at the wallstand.
• The overhead tube crane moves down and align with the wallstand.
• The overhead tube crane follows the wallstand to stay aligned.
• The overhead tube crane turns on the collimator light when the correct height is reached.
Adjust the position.
• The mode stays activated.
• The SID value is changed to the new distance (if selected distance toward the wallstand is
changed).
OMNERA 500A
Select a Wall Flexible mode examination.
See Fig. 4-31.
1000-095-071-EN Rev. 2.1
Page 114
Operator´s Manual