Canon OMNERA 500A Operator's Manual page 120

Hide thumbs Also See for OMNERA 500A:
Table of Contents

Advertisement

Operating the System
X-ray Table Examination Procedures
Rotate the x-ray tube in order for the left edge of the collimator light field to indicate the left
limit for the composite image.
Press the button (F2) to set the left limit, see Fig. 4-28.
• The button turns green to indicate that the limit is set.
• Data is present on the right side of the display.
• When both limits have been defined, the total length of the composite image (C), see
Fig. 4-29, and the number of exposures (D) are shown.
• The servo mode indication light is fixed.
Modify the exposure settings if necessary.
Activate the exposure button, and keep it activated during the procedure.
• The starting position is always at the left edge of the composite image. Activate the
exposure button and keep it activated until the exposure procedure is completed. The
system moves to the correct starting position.
• When the system is in the correct position for the first image, the first image is captured.
• After exposure, the system moves to the next correct position and the second image is
captured.
This is repeated until all images for the composite image has been captured.
• The system beeps when the sequence is finished.
OMNERA 500A
110
3
exp
K
Fig. 4-29 Left and Right Limits Set
1000-095-071-EN Rev. 2.1
Page 112
C
cm
D
Operator´s Manual