Oscilloscope Method - Siemens MOBILETT XP Digital Installation And Startup

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

44
Fig. 18: kV measurement
X
Set the following exposure parameters and release an exposure after each setting:
kV meter
Selection
DC voltage
* The measuring inaccuracy of the measuring instrument used must be subtracted from
the specified limit values.
P
Record the measured values.

Oscilloscope method:

"KVS" measuring point for the nominal value and "KV" measuring point for the actual
value. The measuring ratio is 30 kV/V.
Remove the top cover and place on a table or similar surface. All cables remain con-
nected. Measuring points on CPU D916 (see troubleshooting instructions).
Connect the oscilloscope to the measuring points of the CPU D916 TP KV, KVS, and
GND (also see the troubleshooting instructions). Oscilloscope setting: Channel 1 = TP
KV 1V/div, channel 2 TP KVS 2V/div, trigger channel 2, trigger stage 2.5 V, 2V/div, 50
ms/div
MOBILETT XP Digital
Exposure parameters
52 kV
50 mAs
81 kV
20 mAs
133 kV
12.5 mAs
SPR8-230.814.30.03.02
11.05
Limit values*
49.0 - 55.0 kV
77.0 - 85.0 kV
126.4 - 139.6 kV
Page 44 of 70
CS SD 24
DHHS
0
Siemens AG
Medical Solutions