DHHS
•
Determination for the following exposure parameters (product of tube current x expo-
sure time):
X
* The measuring inaccuracy of the measuring instrument must be subtracted from the
specified limit values.
P
Calculate and record the mAs values.
mAs meter method:
Use of the mAs jumper on PCB D907 directly at the single tank.
•
Remove cover from single tank and connect mAs meter to D907
•
Follow the same procedure as described in "Oscilloscope method".
For more information on the measuring technique, refer to the user's manual for the
mAs meter.
Fig. 20: mAs jumper D907
Siemens AG
Medical Solutions
Control unit setting
kV
40
81
133
SPR8-230.814.30.03.02
11.05
Allowable mAs value*
mAs
5
2
10
Page 47 of 70
CS SD 24
4,7 - 5,3
1,9 - 2,1
9,5 - 10,5
(1/Fig. 20 / p.
MOBILETT XP Digital
47
0
47).