DHHS
Determination for the following exposure parameters (product of tube current x expo-
sure time):
X
* The measuring inaccuracy of the measuring instrument must be subtracted from the
specified limit values.
P
Calculate and record the mAs values.

mAs meter method:

Use of the mAs jumper on PCB D907 directly at the single tank.
Remove cover from single tank and connect mAs meter to D907
Follow the same procedure as described in "Oscilloscope method".
For more information on the measuring technique, refer to the user's manual for the
mAs meter.
Fig. 20: mAs jumper D907
Siemens AG
Medical Solutions
Control unit setting
kV
40
81
133
SPR8-230.814.30.03.02
11.05
Allowable mAs value*
mAs
5
2
10
Page 47 of 70
CS SD 24
4,7 - 5,3
1,9 - 2,1
9,5 - 10,5
(1/Fig. 20 / p.
MOBILETT XP Digital
47
0
47).