Checks with high voltage

Oscilloscope diagrams

MAMMOMAT 300 in mAs mode, see Figure 2-Figure 5.
MAMMOMAT 300 with tube current reduction switched on, Iontomat mode
with Bucky, see Figure 6.
MAMMOMAT 300 in Iontomat mode with tube current reduction switched
off, see Figure 7.
NOTE!
Record the measured kV and mAs value in the measuring protocol page 21-3.
Siemens-Elema AB
Solna, Sweden
Tube P49 Mo,
molybdenum anode 0.3
30 kV 20 mAs (P = 3.75 kW)
Figure 2
30 kV
1 V/T (1 V = 5 kV)
125 mA
1 V/T (1 V = 40 mA)
20 ms/T
SPS-UD
RX B7-120.033.01.10.02
05.98
11-3
MAMMOMAT 300
65 02 186