56
Fig. 28: "Control area" label
Reports
•
Separate the completed 'Start-up report' (chapter 9 of these instructions) from these in-
structions and file it in the system binder, Register 9, "Certificates."
•
Separate the completed DHHS report (chapter 10 of these instructions) from
these instructions and file it in the system binder, Register 9, "Certificates."
•
Separate the completed protective conductor resistance report, leakage current report,
and patient leakage current report (chapter 11 of these instructions) from these instruc-
tions and file them in the system binder, Register 9, "Certificates."
•
Complete the "Installation Protocol" included with the unit and send it (by fax) to the ad-
dress listed in the protocol.
MOBILETT XP Digital
SPR8-230.814.30.03.02
11.05
Remaining work
Page 56 of 70
CS SD 24
0
Siemens AG
Medical Solutions