Fig. 15: Holder
Collimator function
•
P
Completely open the collimator.
•
Check the functioning of the light keys on the collimator.
•
Check the manual collimations with the light on.
•
The light switches off after approx. 20 sec.
•
Test the collimator light key on the control panel
MOBILETT XP Digital
SPR8-230.814.30.03.02
11.05
(1/Fig. 13 / p.
32).