Start-up
Options
DAP
P
Measuring device for dose area product (Dose Area Product).
The DAP measuring device is used to measure the kerma area product (kerma = kinetic
energy released in matter) during an X-ray examination. It is primarily used to record the
radiation dose a patient is exposed to.
The DAP is calibrated at the factory:
•
Press the test button on the DAP display, a value between 80-120 µGym
tion 0.1 µGym
tion 0.01 µGym
If the display is incorrect, the measuring device must be calibrated. See the
"Replacement of Parts" instructions, SPR8-230.841.30...).
•
X
Check the function of the DAP display by releasing an exposure. The DAP display
shows a measurement value depending on the object in the beam path.
Remote control
•
P
Adhere the remote control holder to the front right or left side
•
Switch on the MOBILETT XP Digital.
•
X
Check the "collimator light" and "exposure release" functions of the remote control.
Siemens AG
Medical Solutions
2
) or in the case of a high-resolution chamber, 8-12 µGym
2
) must be subsequently displayed.
SPR8-230.814.30.03.02
11.05
(16/15/Fig. 3 / p.
MOBILETT XP Digital
Page 41 of 70
CS SD 24
41
0
0
2
(DAP resolu-
2
(DAP resolu-
0
12).