UV Ignition Failed

The UV lamp failed to ignite.
The processor monitors the UV lamp current during the ignition cycle. If the
lamp current does not rise above the lower limit within 2 – 5 seconds, the
error message is generated.
• Lamp not connected.
• Defective UV lamp or non-Agilent lamp.
• Defective detector main board.
• Defective power supply.
1100 Series DAD and MWD Reference Manual
Probable Causes
Suggested Actions
Ensure the lamp is connected.
Exchange the UV lamp.
Exchange the detector main board.
Exchange the power supply.
Troubleshooting and Test Functions