The fittings at the flow cell body are special types for
low dead volumes and not compatible with other
fittings.
When retightening the fittings, make sure that they
are carefully tightened (handtight plus 1/4 turn with
a wrench). Otherwise damage of the flow cell body
or blockage may result.
3 To replace the inlet capillary, use a 4-mm wrench for the
fitting.
1100 Series DAD and MWD Reference Manual
2 After replacing the outlet capillary, fix it handtight first.
4 Unscrew the cell body from the heat exchanger and the
Inlet capillary
Repairing the Detector
Then do a 1/4 turn with a 4-mm wrench.
heat exchanger from the clamp unit.
Outlet capillary