Repairing the Detector

Repairing the Standard or Semi-Micro Flow Cell

When required
Tools required
Parts required
The gaskets used in the standard and semi-micro flow cell are different.
N O T E
Preparations for this procedure:
Turn off the flow.
Remove the front cover.
Remove the flow cell, see
page 105.
108
If the flow cell needs repair due to leaks or contaminations (reduced light
throughput)
Two 1/4 inch wrenches for capillary connections
hexagonal key 4 mm
Tooth picks
For parts, see
Cell"
on page 188
"Exchanging a Flow Cell"
"Standard Flow Cell"
on page 186 and
1
Use a 4 mm hex key to unscrew the window assembly and
remove the gasket from the cell body.
on
1100 Series DAD and MWD Reference Manual
"Semi-Micro Flow
gasket
windows
assembly