8. Before calling the service technician (Warning)

8.3 Warning

module id message id
10001
10002
10003
Invalid image was captured.
Failed to acquire Retinal Camera Info. within the
1002
specified time.
No Validity for Current Study
Photography frequency exceeded the limit.
1009
Please wait.
Failed to connect to EOS.
1000
Check the cable.
Check the power supply.
Failed to control EOS digital camera.
Follow the instructions as follows:
1. Turn off LCD monitor on the back if it's on.
2. Close the setting menu if it's displayed.
3. Turn the retinal camera to the eye front image
1001
mode if it's is in retinal image mode.
4. Release Shutter Release button of the retinal
camera if it's held down.
When all the items above are cleared, press Retry
button.
Auto power off option is enabled on EOS.
1004
Select Off for Auto power off.
1005
Communications with EOS was disconnected.
EOS battery is almost empty.
1006
Supply power with AC adapter.
EOS setting has been changed.
1007
Reset the setting?
1
Selected study will be deleted.
3
No Storage Selected
4
Character '\' cannot be used for Patient ID.
7
Character '\' or '=' cannot be used for Patient Name. Delete the backslash (\) or equal sign (=).
Only alphanumeric characters can be used for
8
Patient Name.
The study being transferred is selected. Failed to
203
transfer.
The study being transferred is selected. Failed to
204
delete the study.
Study Logs screen cannot display all logs. Number
of study logs to be displayed exceeds the
11004
maximum.
Change the search conditions.
Message
-116-
Detail
Do not press the shutter button of the digital
camera.
Check that the computer and digital camera are
connected correctly.
Turn the equipment's power off and back on,
and take a test image.
Images cannot be saved since images were taken
continuously over a short period of time.
Wait for a few moments, and then take the images.
Check that the computer and digital camera are
connected correctly.
Check that the digital camera is turned on.
Perform operation based on the message details.
Refer to the operating instructions of the digital
camera, and set the digital camera.
Check that the computer and digital camera are
connected correctly.
Check that the digital camera is turned on.
Supply the power of the digital camera using the AC
adapter.
To return to the original setting, click [YES];
otherwise, click [NO].
To delete a study, click [OK]; to cancel the deletion,
click [Cancel].
Select the storage, and then transfer the image.
Delete the backslash (\).
Delete the characters which are not single-byte
alphanumerics or symbols.
No operations can be performed for studies being
transferred so wait until the transfer is completed.
Change the search conditions (see page 32).