BIOSHIELD™ Commercial UV Disinfection System Installation and User's Guide
MAIN UV System Text Display
The "MAIN UV System Text Display" will appear after initialization. On this screen
you will see: Total System Operating Hours, Input Voltage, and UV Intensity. Total
Hours may be reset and the UV scaling adjusted via the "SETUP" menu. High and
Low Input Voltage (95 – 140 VAC or 210 – 260 VAC depending on the specific
unit's control voltage requirements) are preset at the factory and may not be
changed.
Press "ENTER" to move to the "SETUP" screen or "RIGHT" to move to the
"STATUS" menu.
STATUS Menu Text Display
Four lines of text will appear on the STATUS Menu Text Display. Use the "Right"
key to advance the selection arrow (">") through the options of the STATUS Menu.
Press the "Enter" key to select the desired option. Use the "Left" key to return to
the "MAIN UV System Text Display".
After selecting the desired STATUS Menu option, a brief description will appear
on the Text Display. Press the "Left" key to return to the STATUS MENU. Failing
to press any "Function Keys" within a couple minutes will automatically revert the
Text Display back to the MAIN UV System Screen.
STATUS Menu Text Display Options
Lamp Status
Displays the current On/Off status of lamps 1 through X. 0 = OFF, 1 = ON. For
example, a "four lamp" UV system's lamps are identified on the bottom of the
screen in ascending order from right > left. The display example shown here
indicates all lamps are in operation, except for lamp 3.
Lamp Hours
Displays the current "total" operating hours of each lamp, six per screen.
Each lamp's status is checked once an hour and "Total" operating hours are
incremented by a factor of one. Here, all four lamps show one hour of operation.
UV% Lo Setting
Displays the active UV low set-point as well as the actual UV % intensity. XXX shown here
is an operator editable UV% parameter accessible through the SETUP screens. YYY is the
actual UV % intensity reading taken from the sensor probe positioned on the UV vessel.
The actual UV % intensity (YYY) is updated continuously.