Replenishment Calibration Check Procedure

The Replenishment Pumps are preset at the factory so that they will pump proportional volumes of developer and
fixer replenishment solutions. The replenishment rate for the fixer should always remain 15 to 20% greater than the
replenishment rate for the developer. When checking the replenishment rate, always check the replenishment rates
for both the developer and the fixer, but make any adjustment using the developer replenishment solution.
For the recommended replenishment rates for developer and fixer, see the replenishment rate information sheet,
Publication No. 1C0578, provided in the front pocket of your publications binder.
[1] To enter the Replenishment Calibration Check Mode:
• Check that the processor is on (CB1 in the "1" position).
• Remove the Top Cover.
Note
When the Top Cover is removed, the Safety Interlock is enabled. You will not see or hear any of the individual
components of the processor running.
[2] Insert the Developer Replenishment Check Tube into a Graduated Cylinder.
Note
The Graduated Cylinder must be capable of containing at least 500 mL of solution. Use a 500 mL Graduated
Cylinder TL-1435.
Figure 10 Checking the Developer Replenishment Rate
Fixer
Replenishment
Check Tube
981089 – September 1995
Developer
Replenishment
Check Tube
Replenishment Solutions
Feed Tray
Graduated Cylinder
H130_0021BCB
H130_0021BA
15