Temperature Settings; Calibrating Temperature Sensor - YaliTech LAQUAact-PC110 Instruction Manual

Handheld ph/orp/conductivity meter
Table of Contents

Advertisement

Temperature settings

● Calibrating temperature sensor
The temperature sensor or temperature compensation electrode in the combination
electrode has ±1°C accuracy without calibration. You can use a known temperature
solution to calibrate the temperature sensor.
1. Insert the temperature connector into the jack of temperature connector (T) on
the instrument.
2. Immerse the electrode into the solution until the temperature sensor is
immersed.
3. Press the
connected to the temperature sensor to calibrate is active.
4. Press the
then press the
5. Make sure that "ATC" is displayed, and press the
The temperature setting screen is displayed.
6. Enter the set temperature and press the
The temperature sensor is calibrated.
key to enter the setting mode when the channel which is
keys to select "TEMP" (temperature calibration setting) and
key.
key

Temperature settings

key
key.
key to confirm.
77