How to resolve errors or troubles

This section describes the causes of typical problems and the actions to be taken,
including questions frequently asked by customers. Check these before contacting us.

When an error message appears

If "ERR No.00XX" is displayed while you are using the instrument, check the error in the
error list below, and check the cause and action to be taken.
ERR No.
0001
Memory error
0002
Empty battery level
Asymmetric potential
0004
Electrode sensitivity
0005
Maximum calibration
0006
points exceeded
Cannot identify
0007
standard solution
Calibration interval
0008
0009
0010
Cell constant is out of
0011
● ERR No.0001   Memory error
Data cannot be read from or written to the internal memory.
Cause
The instrument does not start
properly due to noise or other at
power ON.
The defect of the internal IC
Description
Data cannot be read from or written to the internal
memory.
The battery level is empty.
The asymmetric potential of the electrode is out of
error
the range of ±45 mV.
The electrode sensitivity is either 105% or more or
error
85% or less than the theoretical sensitivity.
6th point calibration is attempted.
The instrument cannot identify the standard
solution.
Exceeds the calibration interval setting.
error
Printer error
There is a problem with the printer.
The number of the data saved has exceeded the
Memory full
specified number of items.
Cell constant is out of automatic calculation range.
range
Remove the batteries, disconnect the AC adapter, and
then press the
Contact your dealer for repair.
When an error message appears
Definition of error
How to solve problem
key.