Err No.0008   Calibration Interval Error; Err No.0009   Printer Error; Err No.0010   Memory Full; Err No.0011   Cell Constant Is Out Of Range - YaliTech LAQUAact-PC110 Instruction Manual

Handheld ph/orp/conductivity meter
Table of Contents

Advertisement

When an error message appears
● ERR No.0008   Calibration interval error
・ More than the set number of days has elapsed since calibration was last performed.
・ The calibration interval alarm is "ON" and calibration has not been performed.
Cause
Calibration has not been
performed for the set number of
days of the calibration interval or
longer.
The calibration interval alarm is
"ON" and calibration has not
been performed.
● ERR No.0009   Printer error
An error occurred during printer communication.
Cause
There is a problem with the
printer unit connection.
The defect of the printer
● ERR No.0010   Memory full
Attempted to save more than the specified number of items.
Cause
Saving more than specified
number of items.
● ERR No.0011   Cell constant is out of range
Cell constant is out of setting range.
Cause
End of cell life
There is a problem with the
standard solution
104
How to solve problem
Perform calibration.
Perform calibration.
How to solve problem
Check the printer connection, and connect the instrument
and printer again.
Consult your dealer.
How to solve problem
The maximum number of savable items of data is 1000.
Copy or transfer necessary data to a PC and delete the
data from the memory ("Deleting all saved data" (page
68)).
How to solve problem
Replace the conductivity cell.
Use new standard solution.