Measurement setting
● Using the calibration interval alarm (default: OFF)
If calibration has not been performed for set period of time after last calibration is
performed, "ERR No. 0008" is displayed to prevent forgetting to perform calibration.
When the error is displayed, performing calibration clears the error.
1. Press the
main screen.
2. Press the
3. Press the
press the
4. Enter the setting period of the calibration interval alarm and then press the
key.
The setting range is 1 day to 400 days.
To return to the measurement mode, press the