● Conductivity standard values at various temperatures
Temp.
(°C)
84.00 (S/cm)
0
64.01
5
65.00
10
67.00
15
68.00
16
70.00
17
71.00
18
73.00
19
74.00
20
76.00
21
78.00
22
79.00
23
81.00
24
82.00
25
84.00
26
86.00
27
87.00
28
89.00
29
90.00
30
92.00
31
94.00
Conductivity value at 25°C
1413 (S/cm)
12.88 (mS/cm)
776
896
1020
1147
1173
1199
1225
1251
1278
1305
1332
1359
1386
1413
1440
1467
1494
1521
1548
1575
111.8 (mS/cm)
7.15
65.4
8.22
74.1
9.33
83.2
10.48
92.5
10.72
94.4
10.95
96.3
11.19
98.2
11.43
100.2
11.67
102.1
11.91
104.0
12.15
105.9
12.39
107.9
12.64
109.8
12.88
111.8
13.13
113.8
13.37
115.7
13.62
117.7
13.87
119.7
14.12
121.8
14.37
123.9
Technical note