YaliTech LAQUAact-PC110 Instruction Manual page 87

Handheld ph/orp/conductivity meter
Table of Contents

Advertisement

3. Press the
press the
key.
The electrode status based on the calibration result is displayed. You can change the
display item among "OFS" (asymmetry potential) to SLP (sensitivity) to REP
(repeatability) by pressing the
key
Repeatability display
To return to the measurement mode, press the
Tip
• The asymmetry potential can be displayed only for pH calibration data.
• When multiple point calibration (more than two points) is performed, the results of multiple
point calibration are displayed.
・ Electrode status based on calibration result
Display
Electrode sensitivity: 85% to 105%
GOOD
Good condition.
keys to select "CAL DATA" (display calibration data) and then
keys.
key
key
Asymmetry potential display
Description
key
key and then press the
Measurement setting
key
Sensitivity display
key.
Reference
71