● Changing the calibration method
Select automatic calibration or manual calibration for the conductivity calibration method.
1. Press the
key to enter the setting mode when "COND" icon is shown on the
main screen.
2. Press the
key.
3. Press the
then press the
4. Select "ON" to set the automatic calibration, or select "OFF" to set the manual
calibration. And then press the
key
To return to the measurement mode, press the
Note
When the setting of calibration method (auto or manual) is changed, the cell constant is
initialized. Make sure to enter the cell constant or to perform calibration again.