Emerson Rosemount OCX 8800 Reference Manual page 41

Oxygen and combustibles transmitter
Hide thumbs Also See for Rosemount OCX 8800:
Table of Contents

Advertisement

Figure 2-6:
A.
B.
C.
D.
E.
F.
G.
H.
I.
J.
Reference Manual
Customer-furnished interconnect wiring or cables
Heat shrink tubing, 2 ft long, 1/2 in. size
3/4 NPT hub size, liquid-tight strain relief connector
Heat shrink tubing, 2 ft long, 1/2 in. size
Ferrule, uninsulated
8 twisted pairs 24 AWG, stranded, insulated, tinned copper conductors, 200 °C (392 °F), 300 volts,
with overall braid of 34 AWG tinned copper, 90% coverage and 24 AWG tinned copper, uninsulated
drain wire
See Note.
See Note.
7.0 ft long Teflon tubing, 0.042 in. ID (Cut off drain wire at probe end of shield.)
Stud size #10
Strip wire ends 3/16 in. typical
Install
31