3.
4.
5.
6.
Reference Manual
Note
For easier access, you may remove two screws from the base of the terminal block mounting
(F) and move the terminal block mounting assembly out of the way.
Remove tubes (B, C, and D) from the COe sensor assembly (I), eductor fittings (N
and H), and sensor housing (P).
Unfasten bayonet connector of COe thermocouple (L) and remove thermocouple.
Loosen clamp screw of COe band heater (K) until the heater rotates freely on the
sensor holder.
See
Figure
9-9. Using a straight edge on the sensor holder flat, as shown, matchmark
the upper flange of sensor housing to show the correct alignment of the sensor
holder.
Maintenance and service
223