Maintenance and service
Procedure
1.
2.
3.
4.
240
Apply pipe thread sealant (Loctite #567) to the exposed pipe threads of eductor
elbow (N,
Figure
9-15).
Do not apply sealant to the first turn of the pipe threads.
Screw the sensor holder (J) onto the eductor elbow (N).
With wrenches on eductor elbow (N) and on flats of sensor holder (J), tighten the
sensor holder. Do not allow the eductor elbow to turn.
Tighten the sensor holder (J) to align outside flat with matchmark on sensor housing
flange, as shown in
Figure
9-18.