Foundation Fieldbus
Table 7-3:
Descriptions of PlantWeb Alerts
Alerts
Alarms
NV Memory Fail-
Transducer block:
ure
EEPROM Corrupt
Resource block:
1. Manufactur-
ing
2. Block Integrity
Error
3. NV Integrity
Error
4. ROM Integrity
Error
108
(continued)
What does the
AMS tab
alert indicate?
FF/ Device Alerts
The non-volatile
parameter stor-
age on the CPU
board has be-
come unreliable.
Recommended
action
Description
At startup, wait 2
This alarm gener-
minutes with
ally occurs during
power applied
a startup condi-
and then cycle
tion. Rarely, the
power again or re-
device could be
set the device.
powered down
while a parameter
is being stored to
the non-volatile
memory. The pa-
rameter is then
tagged as bad on
the next power
on, and the mem-
ory contents are
written with de-
fault parameters.
Calibration data
may be lost; reca-
librate the unit. If
the unit does not
recover automati-
cally, the memory
may be faulty; re-
place the elec-
tronics stack. Re-
fer to
Section 8.2
for details.
OCX 8800