Emerson Rosemount OCX 8800 Reference Manual page 290

Oxygen and combustibles transmitter
Hide thumbs Also See for Rosemount OCX 8800:
Table of Contents

Advertisement

SPA with HART Alarm
Procedure
1.
2.
3.
280
See
Figure
B-4. Connect a calibrated ammeter (Fluke Model 87 or equivalent,
accurate to ±0.025%) to the SPA analog output terminals.
Observe polarity.
Connect a 90 to 260 Vac or 22 to 300 Vdc power source to the SPA power terminals.
When connecting an AC power source, use the AC and ACC (AC Common)
terminals. For a DC source, use the AC and ground terminals.
If desired, connect the 4 to 20 mA O
output terminal block to the SPA input terminals.
The OCX must be operational to transmit the O
Note
The O
signal is not required for SPA configuration or calibration. The OCX interface will allow
2
you to observe the O
signal level when the SPA configuration procedure is completed.
2
Observe the front panel of the SPA,
signal wires from the transmitter analog
2
signal. Observe polarity.
2
Figure
B-5:
OCX 8800

Advertisement

Table of Contents
loading

Table of Contents