Table 7-3:
Descriptions of PlantWeb Alerts
Alerts
Alarms
NV Writes Defer-
Resource block:
red
1. NV Writes De-
ferred
High Electronics
Transducer block:
Temperature
High Electronic
Temperature
ADC Failure
Transducer block:
1. ADC Timeout
Error
2. ADC Refer-
ence Error
Reference Manual
(continued)
What does the
AMS tab
alert indicate?
FF/Device Alerts
Device Alerts
This alert indi-
cates that the
electronics tem-
perature has ex-
ceeded 80 °C
(176 °F). The de-
vice will cease to
operate reliably
beyond 85 °C
(185 °F).
Device Alerts
This alert indi-
cates faulty oper-
ation of the de-
vice electronics.
Foundation Fieldbus
Recommended
action
Description
A high number of
writes has been
detected to non-
volatile memory.
To prevent pre-
mature failure of
the memory, the
write operations
have been defer-
red. The data is
saved on a 6 hour
cycle. This condi-
tion usually exists
because a pro-
gram has been
written that
writes to function
block parameters
not normally ex-
pected to be writ-
ten to on a cyclic
basis.
Evaluate mount-
The device may
ing location and
require special
environment.
mounting consid-
erations if instal-
led in a very hot
location.
Cycle power or re-
The analog to dig-
place the elec-
ital converter
tronics stack.
(ADC) is continu-
ously monitored
by the device for
correct operation.
Refer to
Section 8.2
tails.
for de-
109