Operation
February 2020
Figure 6-3: Rosemount Radar Master Plus - Overview Screen
6.3.2
View measurement data in AMS Device Manager and
handheld communicator
Current measurement data of the primary variables are presented on the Overview screen.
To view all current measurement values, do the following:
Procedure
1. Select Service Tools → Variables.
2. Select Mapped Variable, Process, Device, or Signal Quality.
6.3.3
Interpret measurement status
A "Good" or "Bad" status next to a value is an indication of the reliability or integrity of the
data being received, not an indication of whether or not the value is within the configured
upper or lower ranges. A value that triggers an alert, such as a high or low temperature
indication, will change the overall status of the device, but the measurement might still be
indicated as "Good" if the reliability of the data is good.