Emerson Rosemount 5408 Reference Manual page 170

Level transmitters, non-contacting radar with hart protocol
Hide thumbs Also See for Rosemount 5408:
Table of Contents

Advertisement

Specifications and reference data
February 2020
Analog signal on alarm
The transmitter automatically and continuously performs self-diagnostic routines. If a
failure or a measurement error is detected, the analog signal will be driven offscale to alert
the user. High or low failure mode is user-configurable.
Table A-4: Signal on Alarm
Standard
Rosemount standard
NAMUR NE43
Analog saturation levels
The transmitter will drive the output to high or low saturation values if measurement goes
outside the 4-20 mA range values.
Table A-5: Saturation Levels
Standard
Rosemount standard (default and option code
C8)
NAMUR NE43 (option code C4 and C5)
A.2.4
Diagnostics
Alerts
The transmitter is compliant with NAMUR NE 107 Field Diagnostics for standardized
device diagnostic information.
Tools and logging in Rosemount Radar Master Plus
Echo curve
Measurement and alert log
Rosemount Radar Master Plus, embedded in Instrument Inspector, enables easy and
powerful troubleshooting with the echo curve tool as well as the measurement and alert
log.
The measurement and alert log holds records of the last seven days of level readings and
echo curve profiles, as well as the 50 last alert events. The logs can be transferred from the
transmitter's internal memory to a local computer and be presented in a graphical time
line, enabling analysis of historical behaviors.
Smart Diagnostics Suite
Signal Quality
Metrics
170
High
≥ 21.75 mA (default)
≥ 22.50 mA (option code C4)
Diagnostics package that monitors the relations between surface,
noise, and threshold. The function can be used to detect abnormal
conditions in the process such as antenna contamination or sudden loss
Low
≤ 3.75 mA (option code C8)
≤ 3.6 mA (option code C5)
High
Low
20.8 mA
3.9 mA
20.5 mA
3.8 mA
Reference Manual
00809-0100-4408
Reference Manual

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount 5408:sis

Table of Contents