Service and troubleshooting
February 2020
7.4.6
Save a backup file of the device configuration
When configuration is finished, it is recommended to store the device configuration in a
backup file for future reference using Rosemount Radar Master Plus. A backup of the
device configuration will be saved to file as well as a configuration report (optional).
The backup file may be useful to:
Restore the configuration of the transmitter.
Install another transmitter in a similar tank.
Troubleshoot the transmitter.
Procedure
1. In Rosemount Radar Master Plus, under Service Tools, select Maintenance →
2. Select Save Configuration.
3. Type your desired file name.
4. Browse to the desired directory, and then select Save.
5. Optional: Select the Create and save report (.pdf) check box.
6. Select Save.
7.4.7
Download configuration from file to device
Procedure
1. In Rosemount Radar Master Plus, under Service Tools, select Maintenance →
2. Select Restore Configuration.
3. Browse to the backup file and select Open.
7.4.8
Restore to default settings
Restore to default settings using Rosemount Radar Master
Plus
This function restores the transmitter to default settings (user configuration is
overwritten).
Prerequisites
Before restoring the transmitter to default settings, it is recommended to backup the
device configuration. The backup file can be used to restore configuration at a later stage.
Procedure
1. Under Service Tools, select Maintenance → Reset/Restore.
2. Select Restore Default Settings.
128
Backup.
Backup .