Operation
Model 8901B
(50R) nominal input impedance (LO Input and LO Output). Two jumpers are provided to allow
the internal LO signal to be routed to the LO input connector when no external LO is in use.
The Option Series 030 instruments are provided with an LO input connector, an LO output
connector and
a
remotely controllable switch which selects between them. If an external
LO
is
connected but the signal is not in use, the switch routes the external LO signal back out the LO
output connector. [See High Selectivity Measurements.]
Do not apply reverse power into the
u)
OUTPUT
or damage to the
instrument may result. Do not apply greater than 40 Vdc or
+5
dBm
of
RFpower into the LO
INPUT
or damage to the instrument may result.
16. CALIBRATION OUTPUT RF PWR and A M F M produces a rear-panel output for
RF
Power
and AM and FM calibrators. This output is supplied on Option 001 instruments instead of the
standard front-panel connection. Refer to both CALIBRATION RF POWER OUTPUT and
CALIBRATION A M F M OUTPUT referenced from Figure 3-1.
3-10