Model 8901B
Adjustments
NOTE
If the frequency of the power reference i s already within the proper limits,
do not attempt to adjust the frequency to exactly 50 MHz. To
do so may
unnecessarily affect the level.
4.
Reinstall the A32 assembly. (The thumb screw need only be finger tight.)
5. Connect the input of the sensor module (or a compatible power sensor connected
to the
6. On the Modulation Analyzer, press ZERO and wait for the power sensor to zero.
7. On the Modulation Analyzer, press
RF
POWER. Press
RATIO
and check that
LOG/LIN
is in
the linear mode. (The display should read between 99.9 and 100.1%
REL.)
8. Connect the sensor module to the
CALIBRATION
RF
POWER OUTPUT of the Modulation
Analyzer. Press ZERO and allow the instrument to zero. Adjust A32R4
(LEVEL ADJ)
for a
reading between 99.9 and 100.1%
REL. If
the adjustment runs out of range, replace A32R5 with
75000
and readjust A32R4.
Modulation Analyzer's SENSOR input)
to
the output of the external power reference.
Adjustment 18
5-29