HP 8901B Operation And Calibration Manual page 82

Hide thumbs Also See for 8901B:
Table of Contents

Advertisement

1-18
RF
Power
Meter
RF Power Meter
RF Power
Calibration
Error 7 5
Note
1.
2.
3.
The HP 8901B provides you with the capability and accuracy of the
best power meters. With a single keystroke, the HP 8901B autoranges
to the input signal, measures its power, compensates for power
sensor flatness, and displays the results in the units of your choice.
Adjust Your Signal Generator
RF Frequency
....................
100 MHz
Amplitude
.......................
0.0 dBm
Modulation
......................
Off
Like all power meters, the HP 89018 must first be calibrated. An
internal
RF
Power Calibrator enables you to calibrate the Sensor
Module and the HP 8901B's Power Meter to an internal 50 MHz,
1 .OO mW reference signal. The following calibration procedure must
be performed only one time to calibrate your Sensor Module.
This
means you do not need to perform this procedure again as long
as you are using the same Sensor Module. To ensure the highest
possible measurement accuracy, allow the HP 8901B to warm up for
at least thirty minutes before performing the calibration procedure.
Press the
RF POWER
key on your HP 8901B.
If the HP 8901B is displaying Error 15, it has not yet been configured
to make RF Power measurements. (Refer to Appendix B and perform the
Calibration Factor Entry procedure before going on with this procedure.)
The calibration factors for the Sensor Module
you
are using must be
stored in the HP 8901B in order for i t to make calibrated, automatic
RF Power measurements. If
you
are not certain that the correct
calibration factors are stored in the instrument, refer to Appendix B.
Connect the HP 11722A Sensor Module to the RF POWER
OUTPUT port on the
HP
8901B. This port outputs the 1.00 mW
reference signal. (Disregard Error 11 when it is displayed.)
Press the ZERO key. Wait approximately 8 seconds for the zeroed
sensor reading to be displayed briefly.
The
HP 8901B is correcting
for any dc offset error inherent within the measurement system.