Store/Recall
FUNCTIONS
STORE, RECALL
DESCRIPTION
The Modulation Analyzer can store a complete instrument state (including all front-panel functions and
most Special b c t i o n s ) in any of its eight storage registers. The stored data is retained in non-volatile
memory when the instrument is off even when the power cord is removed. Any of these stored states
can then be re-initiated.
PROCEDURE
To store the current instrument state, select STORE (Blue Key,
7).
Then key in a register number
(1 through 8).
To recall a previously-stored instrument state, select RECALL (Blue Key, 8). Then key in the desired
register number (1 through 8).
HP-IB PROGRAM CODES
RECALL = RC
STORE
=
TR
COMMENTS
Regarding storage of RF power calibration factors:
0
The current reference calibration factor and the frequency/calibration-fac-x pair can be stored.
0
Manual calibration factors can be stored.
0
The table of calibration factors cannot be stored with this function, but two tables of calibration
factors can be separately stored in the instrument's non-volatile memory, (Refer to
R F Power
under "Comments" on page
3-79.)
Store/Recall
3-223