HP 8901B Operation And Calibration Manual page 180

Hide thumbs Also See for 8901B:
Table of Contents

Advertisement

Operation
Model
8901B
Abort
Message Check (cont'd)
Description
Check that all the Modulation Ana-
lyzer's HP-IB annunciators are off.
Send the serial-poll-enable bus com-
mand (SPE) to place the Modula-
tion Analyzer in serial-poll mode.
On the Modulation Analyzer, key in
61.3 and then press the SPCL key.
The display should show
1 .O.
This
indicates the Modulation Analyzer is in
serial-poll mode (indicated by the
"1").
Send the Abort message, re-
moving the Modulation Ana-
lyzer from serial-poll mode.
Check that the Modulation Ana-
lyzer's display shows 0.0. This in-
dicates the Modulation Analyzer
properly left serial-poll mode upon
receiving the Abort message.
I
3-62