Operation
Model
8901B
Error
Message
Code
20
Entered value out of range.
Error 12
Error 12 is enabled by Special finction 15.
Corrective Action
Re-enter new value.
Error 13
21
22
23
Error 13 usually only occurs when the power sensor is disconnected during an
F W
power measurement.
Invalid key sequence.
Invalid Special Function prefix.
Invalid Special Function suffix.
Error 14
Invalid HP-IB code.
End of
RF Power calibration factor table
reached
Error 14 will not occur when using a sensor module (such as the H P 11722A) or a power sensor with
an
FtF
switch controlled by voltages accessed from the rear panel. (Refer
to
Remote Control RF Switch
on page 3-109.)
Re-enter correct HP-18 code.
Check number
of
entries.
Error 15
Refer to R F Power on page 3-74 for entry and edit information concerning calibration factors.
Error 17
Several conditions can cause this error:
0
IF
level too high for available range. Either Range Hold
was
selected or one of the measurement
ranges (Special hnctions 1 through 10)
w a s
user selected. Select AUTOMATIC OPERATION
and try the measurement again.
0
IF level too low when making a reference measurement. The reference
is below 1.25V. Increase
the source level to between 1.25V and 2.5V.
0
Either a ratio measurement was attempted with an inadequate reference measurement or the
reference was not measured before the ratio measurement was attempted. Take a reference
measurement with the reference level between 1.25V and 2.5V.
Error 96
Error 96 corresponds to a display of two dashes
(- -).
Error 96 can occur when no signal is applied
or when Error 03 is disabled and the instrument is manually tuned where no signal is found. Error 96
(- -)
turns off MODULATION OUTPUT/AUDIO INPUT.
Check for compatibility of functions selected.
Re-enter correct Special Function Code.
Re-enter correct Special function Code.
Error 20
When this error occurs during single-sideband noise measurements, the bandwidth entry was out-
of-range. Usually it is unnecessary to enter a bandwidth for noise measurements. This bandwidth is
established at the factory.
3-250
Error Message Summary